FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2000576
·
Received February 1, 2011
Report
- Report Number
- 2531779-2011-00577
- Event Type
- Malfunction
- Date Received
- February 1, 2011
- Report Date
- January 4, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED INTACT WITH NO LIFTING OR PEELING OBSERVED, THE UP/DOWN KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS DURING TESTING, AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION FOUND UNDER ALL THE KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE UP AND DOWN ARROW KEYPAD BUTTONS REPORTEDLY REQUIRE EXCESSIVE FORCE AND NEED TO BE PRESSED SEVERAL TIMES FOR A RESPONSE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.