FDA Adverse Event Injury Summary report: N

PRECISION

MDR report key: 2000476 · Received February 16, 2011

Report

Report Number
3006630150-2011-00219
Event Type
Injury
Date Received
February 16, 2011
Date of Event
January 26, 2011
Report Date
January 26, 2011
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P030017
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE EXPLANTED IPG PASSED VISUAL, PHOTOGRAPHIC IMAGING AND PERFORMANCE TESTS DONE. THE DEPLETION RATE WITH THE STIMULATION OFF WAS HIGHER THAN EXPECTED. IT WAS DETERMINED THIS SLIGHTLY HIGHER THAN NORMAL CURRENT DRAIN WAS FROM THE ANALOG/DIGITAL IC SECTION OF THE IPG ELECTRONICS. IT COULD NOT BE DETERMINED CONCLUSIVELY, WHICH IC IS THE ROOT CAUSE OF THE FAILURE AS BOTH COMPONENTS ARE COVERED IN GLOP TOP WHICH MAKES TEST POINTS INACCESSIBLE, AND TROUBLESHOOTING TO SPECIFIC COMPONENT LEVEL IS NOT POSSIBLE.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PT IS EXPERIENCING CHARGING DIFFICULTY. A BSN ENGINEER ANALYZED THE PATIENT'S DATABASE AND CONFIRMED PREMATURE DEPLETION. THE PT UNDERWENT AN IPG REPLACEMENT SURGERY AND IS DOING WELL.

Description of Event or Problem · 1

A REPORT WAS RECEIVED THAT THE PATIENT IS EXPERIENCING CHARGING DIFFICULTY. A BSN ENGINEER ANALYZED THE PATIENT'S DATABASE AND CONFIRMED PREMATURE DEPLETION. THE PATIENT UNDERWENT AN IPG REPLACEMENT SURGERY AND IS DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION SPINAL CORD STIMULATOR LGW BOSTON SCIENTIFIC NEUROMODULATION SC-1110-02 NA

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention