FDA Adverse Event Death Summary report: N

MEDTRONIC ANEURX STENT GRAFT SYSTEM

MDR report key: 2000449 · Received February 18, 2011

Report

Report Number
2953200-2011-00455
Event Type
Death
Date Received
February 18, 2011
Report Date
January 19, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVALUATION: RESULTS: (RUPTURE, DEATH).

Description of Event or Problem · 1

AN ANEURX BIFURCATED STENT GRAFT AND ITS COMPONENTS WERE IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT 72 MONTHS AGO THE ANEURYSM MEASURES 63MM IN DIAMETER AND THE AORTIC NECK DIAMETER HAS INCREASED FROM 22MM TO 27MM IN DIAMETER. IT WAS REPORTED THAT DUE TO A TYPE 2 ENDOLEAK, THE PATIENT HAD A SUCCESSFUL COIL EMBOLIZATION PROCEDURE DONE. A FEW MONTHS LATER, A CT REVEALED THAT THE STENT GRAFT MIGRATED DISTALLY AND A SMALL TYPE I ENDOLEAK WAS PRESENT. THE PHYSICIAN REQUESTED A TALENT AORTIC CUFF AS A COMPASSIONATE USE FOR TREATMENT FOR THIS PATIENT. THE TALENT CUFF WAS SUCCESSFULLY IMPLANTED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED. (REF MFR# 2953200-2004-01124). ADDITIONAL INFORMATION WAS RECEIVED FOR THIS CASE. IT WAS REPORTED THAT THE PATIENT HAD OPEN REPAIR ON 60 MONTHS AFTER THE SECONDARY INTERVENTION DUE TO A RUPTURING ANEURYSM. IT WAS REPORTED THAT THE RUPTURE WAS RELATED TO A SIGNIFICANT TYPE 3 ENDOLEAK SEPARATION OF ANOTHER MANUFACTURER'S STENT GRAFT AND A TALENT EXTENSION CUFF (REF MFR# 2953200-2011-00454). IT WAS NOT POSSIBLE TO PERFORM ENDOVASCULAR REPAIR. IT WAS REPORTED THAT THE PATIENT LATER EXPIRED 18 MONTHS LATER OF UNKNOWN CAUSES. REF. MFR# 2953200-2011-00456).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDTRONIC ANEURX STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA M00K551454

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death