ENDURANT ABDOMINAL STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-00447
- Event Type
- Injury
- Date Received
- February 18, 2011
- Date of Event
- January 20, 2011
- Report Date
- January 20, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR GALWAY
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: (STENT GRAFT MISPLACEMENT, OCCLUSION, PERFORATION). (INADVERTENTLY COVERING THE HYPOGASTRIC ARTERY AND AGGRESSIVE BALLOONING). EVAL, CONCLUSION: (INADVERTENTLY COVERING THE HYPOGASTRIC ARTERY AND AGGRESSIVE BALLOONING).
AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5.2 CM ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS THE AORTIC NECK WAS 2 CM LONG AND MEASURED 24 MM IN DIAMETER BELOW THE RENAL ARTERIES, THEN IT WENT TO 26 AND 28 MM JUST ABOVE THE ANEURYSM. THE ILIAC VESSELS MEASURED 9 MM IN DIAMETER BILATERALLY. THERE WAS NO TORTUOSITY OR CALCIFICATION NOTED. IT WAS REPORTED THAT BOTH HYPOGASTRIC ARTERIES WERE INADVERTENTLY COVERED WITH THE STENT GRAFTS. THE PHYSICIAN RE-WIRED ONE OF THE HYPOGASTRIC ARTERIES AND IMPLANTED A STENT. DURING THE BALLOONING, WHICH WAS TOO AGGRESSIVE AND THE ILIAC ARTERY AT THE ILIAC BIFURCATION WAS PERFORATED BY AN OTHER MANUFACTURER'S BALLOON. THE PHYSICIAN EXTENDED WITH A 16 X 10 ENDURANT ILIAC LIMB, COVERING THE RUPTURED VESSEL. THE PT LOST ABOUT A UNIT OF BLOOD. AT THE END OF THE CASE THERE WAS NO ENDOLEAK AND THE ILIAC WAS PATENT. NO CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE. (REF MFR# 2953200-2011-00448).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDURANT ABDOMINAL STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR GALWAY | NA | V00799478 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention |