FDA Adverse Event Malfunction Summary report: N

FLUSHING PUMP OFP-2 (JP)

MDR report key: 20001888 · Received August 20, 2024

Report

Report Number
9611174-2024-01436
Event Type
Malfunction
Date Received
August 20, 2024
Date of Event
August 9, 2024
Report Date
September 30, 2024
Manufacturer
KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
Product Code
FEQ
PMA / PMN Number
K100899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E2/E3: NO INFORMATION AVAILABLE FOR THE OCCUPATION OF THE INITIAL REPORTER OR WHETHER THEY ARE A HEALTHCARE PROFESSIONAL. THE EVALUATION OF THE EVENT IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE LEGAL MANUFACTURER'S FINAL INVESTIGATION RESULTS AND ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER. REVIEWING THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: THE FLOW RATE CONTROLS BUTTON DID NOT RESPOND, AND THE POWER INDICATOR DID NOT ILLUMINATE. BASED ON THE INVESTIGATION'S RESULTS, IT IS LIKELY THAT THE FOLLOWING LED TO THE MALFUNCTION: THIS EVENT WAS CAUSED BY A COMPONENT FAILURE OF THE CONTROL PANEL. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BUTTON ON THE FLUSHING PUMP DID NOT WORK AS INTENDED. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BUTTON ON THE FLUSHING PUMP DID NOT WORK AS INTENDED. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1209838 FLUSHING PUMP OFP-2 (JP) OLYMPUS FLUSHING PUMP FEQ KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. K10001142

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown