FDA Adverse Event Malfunction Summary report: N

HEART START XL

MDR report key: 2000139 · Received February 18, 2011

Report

Report Number
1218950-2011-00449
Event Type
Malfunction
Date Received
February 18, 2011
Report Date
January 19, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
MKJ
PMA / PMN Number
K001725
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THAT THE BATTERY FAILED TO CHARGE. THERE WAS NO REPORT OF PT INVOLVEMENT. THE UNIT WAS EVALUATED LOCALLY BY A PHILIPS FIELD SERVICE ENGINEER, AND THE FAILURE WAS VERIFIED. REPLACEMENT OF THE POWER SUPPLY RESOLVED THE FAILURE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE BATTERY FAILED TO CHARGE. THERE WAS NO REPORT OF PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEART START XL MKJ PHILIPS HEALTHCARE M4735A