FDA Adverse Event Malfunction Summary report: N

POSEY KEEPSAFE DELUXE

MDR report key: 2000098 · Received February 3, 2011

Report

Report Number
2020362-2011-00032
Event Type
Malfunction
Date Received
February 3, 2011
Report Date
January 5, 2011
Manufacturer
J. T. POSEY COMPANY
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVALUATION: RESULTS: EVALUATION OF THE RETURNED PRODUCT FOUND THAT THE ALARM HAS POWER BUT WHEN IT IS SET ON VOICE ONLY MODE IT DOES NOT SOUND. WHEN IT IS SET ON VOICE AND TONE MODE, IT PLAYS A TONE BUT NO VOICE IS HEARD. THE POWER SWITCH IS STUCK IN THE ON POSITION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE ALARM HAS POWER BUT WHEN THE ALARM IS SET ON VOICE AND TONE AND PRESSURE IS OFF THE PAD THERE IS NO ALARM TONE. CUSTOMER CLAIMS THE ALARM LOOKS NEW AND HAS NO CRACKS ON THE CASE. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POSEY KEEPSAFE DELUXE KMI J. T. POSEY COMPANY 8374 NA

Patients

Seq Age Sex Outcome Treatment
1 NI OVER-THE-MATTRESS SENSOR PAD MODEL 8307, LOT# UNK