FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3100 SYSTEM
MDR report key: 2000049
·
Received February 18, 2011
Report
- Report Number
- 3002158293-2011-00178
- Event Type
- Malfunction
- Date Received
- February 18, 2011
- Date of Event
- January 18, 2011
- Report Date
- February 11, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (MONITOR UNABLE TO CONNECT WITH THE MONITOR) HAS BEEN CONFIRMED. UPON EVAL, PINS IN THE ELECTRODE BELT'S TRUNK CABLE CONNECTOR WERE BENT IS A SWIRL PATTERN. THE CAUSE FOR THE BENT PINS CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY DUE TO THE CONNECTOR BEING FORCED INTO THE MONITOR WHILE THE PINS WERE MISALIGNED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE ELECTRODE BELT IS NO LONGER ABLE TO CONNECT WITH THE MONITOR. THE PT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |