FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 2000032
·
Received February 18, 2011
Report
- Report Number
- 3002158293-2011-00183
- Event Type
- Malfunction
- Date Received
- February 18, 2011
- Date of Event
- January 3, 2011
- Report Date
- February 16, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY WON'T HOLD CHARGE) HAS BEEN CONFIRMED. UPON EVALUATION, THE BATTERY CONNECTOR WAS SHORTED DAMAGING A TRACE ON THE PCA BOARD. THE CAUSE FOR THE SHORTED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY LIQUID INGRESS. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER CHARGER WILL NOT RECOGNIZE ONE OF HER BATTERIES. THE PT WAS ISSUED A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |