FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 2000032 · Received February 18, 2011

Report

Report Number
3002158293-2011-00183
Event Type
Malfunction
Date Received
February 18, 2011
Date of Event
January 3, 2011
Report Date
February 16, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY WON'T HOLD CHARGE) HAS BEEN CONFIRMED. UPON EVALUATION, THE BATTERY CONNECTOR WAS SHORTED DAMAGING A TRACE ON THE PCA BOARD. THE CAUSE FOR THE SHORTED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY LIQUID INGRESS. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER CHARGER WILL NOT RECOGNIZE ONE OF HER BATTERIES. THE PT WAS ISSUED A REPLACEMENT BATTERY PACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 29 YR