FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3100 SYSTEM
MDR report key: 2000007
·
Received February 18, 2011
Report
- Report Number
- 3002158293-2011-00150
- Event Type
- Malfunction
- Date Received
- February 18, 2011
- Date of Event
- September 2, 2010
- Report Date
- February 7, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (LOOSE BELT CONNECTOR) HAS BEEN CONFIRMED. THE ELECTRODE BELT CONNECTOR LOCKING NUT WAS MISSING. THE ROOT CAUSE FOR THE MISSING LOCKING NUT WAS NOT POSITIVELY IDENTIFIED, BUT WAS LIKELY DUE TO EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
THE TERRITORY MANAGER ASSISTING A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S ELECTRODE BELT CONNECTOR IS LOOSE AT THE TOP OF THE MONITOR. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3100 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |