FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 19998384 · Received August 16, 2024

Report

Report Number
3003442380-2024-21555
Event Type
Malfunction
Date Received
August 16, 2024
Date of Event
July 15, 2024
Report Date
October 1, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY THE INVESTIGATION ASSOCIATED WITH RELATED EVENT 1944543 HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE ON MARKET QUALITY REVIEW (OMQR). THE IDENTIFIED MALFUNCTION SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, IS NOT ASSOCIATED WITH A DESIGN OR MANUFACTURING-RELATED COMPLAINT ISSUE. THEREFORE, A DETAILED INVESTIGATION OR INSPECTION OF REFERENCE SAMPLES IS NOT REQUIRED. BATCH REVIEW: LOT 6003303 WAS MANUFACTURED ON 18-SEP-2023, IN LINE 1, WITH A TOTAL OF (B)(4). THE BATCH RECORD WAS REVIEWED TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED, AND NO ISSUES WERE FOUND. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WAS FOUND WITHIN THE DOCUMENTATION. CONCLUSION SUMMARY OF THE RELATED EVENT: DUE TO THE FOLLOWING BATCH RECORD REVIEW YIELDING NO DISCREPANCIES AND NO NON-CONFORMANCE (NC) WAS GENERATED DURING PRODUCTION. DURING MANUFACTURING OF THE CANNULA PART, THE SOFT CANNULA IS KEPT STRAIGHT BY THE INTRODUCER NEEDLE, NO SAMPLES WERE RETURNED FOR INSPECTION. HOWEVER, DURING USE IN GENERAL AND SPECIFICALLY DURING INSERTION MAY ACCIDENTALLY KINK THE SOFT CANNULA. NO FURTHER ACTION IS REQUIRED FOR THIS COMPLAINT, IF USED/UNUSED SAMPLES ARE RECEIVED, APPROPRIATE ACTIONS WILL BE TAKEN.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET CANNULA KINKED EVENT ON 15-JUL-2024. THE EVENT OCCURRED WITHIN 3 HOURS OF INSERTION. THE INSERTION SITE WAS AT THE ABDOMEN AND UPPER BUTTOCKS. THE PATIENT WAS HOSPITALIZED BECAUSE BLOOD GLUCOSE LEVEL WAS 35 MMOL/L WITH THE TRACE OR SMALL AMOUNT OF KETONS AT THE TIME OF HOSPITALIZATION. THEREFORE THE PATIENT WAS TREATED WITH CORRECTION BOLUS VIA PUMP, INTRAVENOUS (IV) OF SALINE AND FLUIDS AND INJECTIONS OF INSULIN VIA MULTIPLE DAILY INJECTIONS (MDI). THE PATIENT HAS NOT BEEN RELEASED. COMPANY DO NOT SEE KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2373153 AUTOSOFT 90 UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1002817 6003303 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 25 YR Male Hospitalization| R