Description of Event or Problem · 1
UPPER MIDDLE PART OF THE HIGH FLOW CIRCUIT MELTED AND STUCK TO BED LINEN. THE HUMIDIFIER DID NOT ALARM; TEMPERATURE READING 28 DEGREES CELSIUS; CIRCUIT FELT HOT TO THE TOUCH (FELT HOTTER THAN 37 DEGREES C). CONDENSATION NOTED BELOW THE MELTED PART OF CIRCUIT; VERY LITTLE CONDENSATION NOTED ABOVE IT. DETACHED HIGH FLOW CIRCUIT FROM THE NSALA CANNULA AND UNPLUGGED POWER PLUG OF NEPTUNE HUMIDIFIER. ADMINISTERED 02 BLOW BY VIA BAG AND MASK; PATIENT SAT SATURATION RECOVERED. THE MELTED CIRCUIT HAD BEEN IN USE FOR APPROXIMATELY 7 DAYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCHATHERM NEPTUNE | HEATED HUMIDIFIER | BTT | TELEFLEX INC | * | * | |
| 2 | HEATED WIRE CIRCUIT | HEATED WIRE CIRCUIT | OFP | TELEFLEX INC HUDSON RCI | * | * | |
| 3 | CONCHA COLUMN | CONCHA COLUMN | BTT | TELEFLEX INC HUDSON RCI | * | * |