ABL800 FLEX
Report
- Report Number
- 3002807968-2024-00047
- Event Type
- Malfunction
- Date Received
- August 15, 2024
- Date of Event
- July 30, 2024
- Report Date
- January 21, 2025
- Manufacturer
- RADIOMETER MEDICAL APS
- Product Code
- CHL
- PMA / PMN Number
- K041874
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IT IS BELIEVED THAT THE ISSUE WAS CAUSED BY A PROBLEM IN THE REFERENCE MEASURING CHAMBER, E.G. DEPOSITION OF A BLOOD CLOT OR PRESENCE OF AN AIR BUBBLE IN FRONT OF THE REFERENCE MEMBRANE.
INITIAL DATE RECEIVED BY MANUFACTURER (AWARENESS DATE) IS 08AUG2024 AND NOT 07AUG2024 AS STATED IN INITIAL REPORT.
ACCORDING TO THE COMPLAINT, THE ABL800 FLEX BLOOD GAS ANALYZER (SERIAL NUMBER: (B)(6) OCCURRED SOME ELECTROLYTES ABNORMAL RESULTS WITHOUT ANY ERROR ON THE SAMPLE. A DOCTOR NOTICED ITS RESULT WAS ABNORMAL THEREFORE IT WASN'T AFFECTING ANY HEALTH HAZARD TO THE PATIENT. CUSTOMER INFORMED OF FOLLOWING MEASUREMENTS: SAMPLE ID: (B)(6), (B)(6) 2024 09:07, NA 170 MQ/L (DISCREPANT VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:07, K 6.4 MQ/L (DISCREPANT VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:07, CA 6.27 MQ/DL (DISCREPANT VALUE). . SAMPLE ID: (B)(6) (B)(6) 2024 -30 09:09, NA 140 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:09, K 5.3 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:09, CA 4.32 MQ/DL (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, NA 140 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, K 5.3 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, CA 4.27 MQ/DL (COMPARISON VALUE). THE CUSTOMER HAS EXPERIENCED HIGH VALUES OF NA, K AND CA ON THE ABL800 FLEX BLOOD GAS ANALYZER (SERIAL NUMBER: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1683923 | ABL800 FLEX | ABL800 FLEX | CHL | RADIOMETER MEDICAL APS | 393-800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |