FDA Adverse Event Malfunction Summary report: N

ABL800 FLEX

MDR report key: 19989707 · Received August 15, 2024

Report

Report Number
3002807968-2024-00047
Event Type
Malfunction
Date Received
August 15, 2024
Date of Event
July 30, 2024
Report Date
January 21, 2025
Manufacturer
RADIOMETER MEDICAL APS
Product Code
CHL
PMA / PMN Number
K041874
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IT IS BELIEVED THAT THE ISSUE WAS CAUSED BY A PROBLEM IN THE REFERENCE MEASURING CHAMBER, E.G. DEPOSITION OF A BLOOD CLOT OR PRESENCE OF AN AIR BUBBLE IN FRONT OF THE REFERENCE MEMBRANE.

Additional Manufacturer Narrative · 0

INITIAL DATE RECEIVED BY MANUFACTURER (AWARENESS DATE) IS 08AUG2024 AND NOT 07AUG2024 AS STATED IN INITIAL REPORT.

Description of Event or Problem · 0

ACCORDING TO THE COMPLAINT, THE ABL800 FLEX BLOOD GAS ANALYZER (SERIAL NUMBER: (B)(6) OCCURRED SOME ELECTROLYTES ABNORMAL RESULTS WITHOUT ANY ERROR ON THE SAMPLE. A DOCTOR NOTICED ITS RESULT WAS ABNORMAL THEREFORE IT WASN'T AFFECTING ANY HEALTH HAZARD TO THE PATIENT. CUSTOMER INFORMED OF FOLLOWING MEASUREMENTS: SAMPLE ID: (B)(6), (B)(6) 2024 09:07, NA 170 MQ/L (DISCREPANT VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:07, K 6.4 MQ/L (DISCREPANT VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:07, CA 6.27 MQ/DL (DISCREPANT VALUE). . SAMPLE ID: (B)(6) (B)(6) 2024 -30 09:09, NA 140 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:09, K 5.3 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:09, CA 4.32 MQ/DL (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, NA 140 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, K 5.3 MQ/L (COMPARISON VALUE). SAMPLE ID: (B)(6), (B)(6) 2024 09:12, CA 4.27 MQ/DL (COMPARISON VALUE). THE CUSTOMER HAS EXPERIENCED HIGH VALUES OF NA, K AND CA ON THE ABL800 FLEX BLOOD GAS ANALYZER (SERIAL NUMBER: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1683923 ABL800 FLEX ABL800 FLEX CHL RADIOMETER MEDICAL APS 393-800

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown