FDA Adverse Event Malfunction Summary report: N

ACUFEX DIRECTOR MRR ANGLED BULLET

MDR report key: 19988555 · Received August 14, 2024

Report

Report Number
1219602-2024-01744
Event Type
Malfunction
Date Received
August 14, 2024
Date of Event
July 20, 2024
Report Date
September 23, 2024
Manufacturer
SMITH & NEPHEW, INC.
Product Code
NBH
UDI-DI
00885556533314
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: INTERNAL COMPLAINT REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

A DEVICE DEFICIENCY WAS NOT IDENTIFIED, AND THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION. INSUFFICIENT PRODUCT IDENTIFICATION INFORMATION WAS PROVIDED AND THUS A MANUFACTURING RECORD REVIEW COULD NOT BE CONDUCTED. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATIONS UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND NO SIMILAR REPORTED EVENTS. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. PLEASE REFER TO THE INSTRUCTIONS FOR USE FOR PRECAUTIONS AND WARNINGS RELATED TO THE USE OF THE DEVICE. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. IF ANY ADDITIONAL INFORMATION RELATED TO THIS EVENT IS RECEIVED IN THE FUTURE, THIS INVESTIGATION WILL BE REOPENED FOR EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ACUFEX DIRECTOR ROOT REPAIR BULLET DID NOT WORK, THE CANNULA OF THE MENISCAL ROOT KIT DID NOT PASS THROUGH THE BULLET. THE PROCEDURE WAS COMPLETED USING A S+N BACK UP DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1895003 ACUFEX DIRECTOR MRR ANGLED BULLET ACCESSORIES,ARTHROSCOPIC NBH SMITH & NEPHEW, INC. UKNOWN 00885556533314

Patients

Seq Age Sex Outcome Treatment
1 NA Female