ZIO AT
Report
- Report Number
- 3007208829-2024-00377
- Event Type
- Malfunction
- Date Received
- August 14, 2024
- Date of Event
- June 2, 2024
- Report Date
- June 20, 2025
- Manufacturer
- IRHYTHM TECHNOLOGIES, INC
- Product Code
- QYX
- UDI-DI
- 00869770000210
- PMA / PMN Number
- K163512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE ZIO AT DEVICE WAS RETURNED TO IRHYTHM, AND THE CLINICAL DATA WAS DOWNLOADED. A REVIEW OF THE CLINICAL DATA FOUND THAT THE PATIENT WORE THE ZIO AT DEVICE FOR THREE DAYS PAST THE 7-DAY PRESCRIBED WEAR PERIOD. IRHYTHM BECAME AWARE OF THE ARRHYTHMIA WHILE PREPARING THE FINAL REPORT AND NOTIFIED THE HCP ON DAY 19. THE INVESTIGATION CONCLUDED THAT THE ALGORITHM DID NOT DETECT THE ARRHYTHMIA DURING WEAR; HOWEVER, IT WAS DETECTED DURING POST-WEAR PROCESSING. THIS EVENT IS BEING REPORTED PER 21CFR 803 AS A PRODUCT PROBLEM /MALFUNCTION. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY IRHYTHM THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED. THESE TERMS ARE INCLUDED IN FORM FDA 3500A AND ARE FIXED TERMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING.
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED IN RESPONSE TO THE FDA NOTIFICATION RECEIVED ON SEPTEMBER 26, 2024, CONCERNING MISSING UDI NUMBERS IN MDR SUBMISSIONS FROM OCTOBER 2023 THROUGH DECEMBER 2024. IN RESPONSE TO THIS ISSUE, IRHYTHM CONDUCTED AN INVESTIGATION AND IDENTIFIED A SYSTEM PROCESSING ERROR, WHICH HAS NOW BEEN RESOLVED. PLEASE SEE THE UPDATE IN SECTION D4.
THE PATIENT EXPERIENCED AN ARRHYTHMIA THAT MET MEDICAL DOCTOR NOTIFICATION (MDN) REQUIREMENTS THAT WAS NOT TRANSMITTED DURING THE WEAR PERIOD. THE INVESTIGATION CONCLUDED THAT THE ALGORITHM DID NOT DETECT THE ARRHYTHMIA DURING WEAR. THE HEALTHCARE PROVIDER (HCP) WAS IMMEDIATELY NOTIFIED, AND IRHYTHM LEARNED THAT THE HCP WAS ALREADY AWARE OF THE PATIENT'S ARRHYTHMIA AND THAT THERE WERE NO DELAYS IN TREATMENT. NO ADVERSE EVENTS, SUCH AS DEATH OR SERIOUS INJURY, ARE KNOWN TO HAVE OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1163184 | ZIO AT | DETECTOR AND ALARM, ARRHYTHMIA | QYX | IRHYTHM TECHNOLOGIES, INC | 00869770000210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Male |