FDA Adverse Event Malfunction Summary report: N

DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 19984588 · Received August 14, 2024

Report

Report Number
3004753838-2024-203678
Event Type
Malfunction
Date Received
August 14, 2024
Date of Event
July 24, 2024
Report Date
December 16, 2024
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
K213919
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

(B)(4). ¿B5: DESCRIBE EVENT OR PROBLEM ¿ ADDITIONAL INFORMATION. ¿D9: DEVICE RETURNED TO MFR - ADDITIONAL INFORMATION. ¿D9: DATE DEVICE RETURNED - ADDITIONAL INFORMATION. ¿H2 TYPE OF FOLLOW UP: ADDITIONAL INFORMATION. ¿H3: DEVICE EVALUATED BY MFG- ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

(B)(4) WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE. 3004753838-2024-203678-01. 3004753838-2024-203678-02.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SIGNAL LOSS OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

PRODUCT HAS BEEN RETURNED AND THE INVESTIGATION IS BEING REVIEWED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED AFTER THE REVIEW IS COMPLETE.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PRODUCT WAS PROVIDED FOR INVESTIGATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. PERFORMANCE DATA WAS REVIEWED. SIGNAL LOSS WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, A TRANSMITTER FAILURE WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE TRANSMITTER FAILURE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2350504 DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9500-161 1724096002

Patients

Seq Age Sex Outcome Treatment
1 64 YR Male