TAPERLOC COMPLETE 133 MP TYPE1 PPS HO 17.0
Report
- Report Number
- 0001825034-2024-02028
- Event Type
- Malfunction
- Date Received
- August 14, 2024
- Date of Event
- July 26, 2024
- Report Date
- December 18, 2024
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KWA
- UDI-DI
- 00880304513570
- PMA / PMN Number
- K110400
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
(B)(4). D10: 51-104150 ITEM NAME TPRLC 133 T1 PPS HO 15X150MM LOT # 7089476. 51-107170 ITEM NAME TPRLC 133 MP TYPE1 PPS HO 17.0 LOT # 7247937. 51-106150 ITEM NAME TPRLC 133 MP TYPE1 PPS SO 15.0 LOT # 6945746. G2: FOREIGN: JAPAN. THE PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET. THE INVESTIGATION IS CURRENTLY IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED. UPDATED: B4, G3, G6, H2, H6 AND H11. VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAS NOT BEEN COMPROMISED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. THE DEVICE EVALUATION FOUND THE PRODUCT TO BE CONFORMING AND STERILITY NOT COMPROMISED. FURTHER, THE EVENT DID NOT CONTRIBUTE TO A SERIOUS INJURY; THEREFORE, THIS WOULD NOT BE CONSIDERED A REPORTABLE EVENT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT WHEN INSPECTING THE PRODUCT, IT WAS IDENTIFIED AS HAVING DAMAGED STERILE PACKAGING. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1045272 | TAPERLOC COMPLETE 133 MP TYPE1 PPS HO 17.0 | PROSTHETIC, HIP | KWA | ZIMMER BIOMET, INC. | N/A | 6402029 | 00880304513570 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | PLEASE SEE H11 |