ENVEO PRO DELIVERY SYSTEM
Report
- Report Number
- 2025587-2024-04586
- Event Type
- Injury
- Date Received
- August 14, 2024
- Date of Event
- August 6, 2024
- Report Date
- September 5, 2024
- Manufacturer
- MEDTRONIC HEART VALVES DIVISION
- Product Code
- NPT
- PMA / PMN Number
- P130021
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IMF CODE MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
UPDATED B.5 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: EVOLUTPRO-26-US, SERIAL/LOT #: (B)(6), UBD: 14-FEB-2026, UDI#: (B)(4) PRODUCT ID: L-ENVPRO-16-US, SERIAL/LOT #: (B)(6), UBD: 2026-MAR-26, UDI#: (B)(4). OTHER CONCOMITANT DEVICE (S) ARE: PRODUCT ID: EVOLUTPRO-23-US, SERIAL/LOT #: (B)(6), UBD: 11-AUG-2024, UDI#: (B)(4). PRODUCT ID: L-ENVPRO-1623-US, SERIAL/LOT #: (B)(6), UBD: 2026-JAN-16, UDI#: (B)(4). PRODUCT ID: ENVPRO-16-US, SERIAL/LOT #: (B)(6), UBD: 2026-FEB-06, UDI#: (B)(4). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
ADDITIONAL INFORMATION WAS RECEIVED THAT A 26 MILLIMETER (MM) VALVE (J217028) WAS ATTEMPTED FIRST AND DISLODGED. A SECOND 23 MM VALVE (F190756) WAS THEN ATTEMPTED. BOTH VALVES WERE EXPLANTED.
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT AFTER THE 23MM VALVE WAS IMPLANTED, THE PATIENT EXPERIENCED HEMODYNAMIC INSTABILITY. BOTH VALVES WERE SUBSEQUENTLY EXPLANTED.
ADDITIONAL INFORMATION WAS RECEIVED THAT A PRE-IMPLANT BALLOON AORTIC VALVULOPLASTY (BAV) WAS PERFORMED. PRIOR TO WITHDRAWING THE DELIVERY CATHETER SYSTEM (DCS), THE NOSE CONE WAS NOT CENTERED WITHIN THE FRAME INFLOW. PER THE PHYSICIAN, THE PATIENT'S AORTIC CALCIFICATION CONTRIBUTED TO THE DISLODGEMENT.
MEDTRONIC RECEIVED INFORMATION THAT DURING IMPLANT OF THIS TRANSCATHETER BIOPROSTHETIC VALVE, NO MISLOAD WAS IDENTIFIED DURING LOADING. DURING THE VALVE IMPLANT, THE NOSE CONE OF THE DELIVERY CATHETER SYSTEM (DCS) DISLODGED THE VALVE. A PROCEDURAL DELAY RESULTED FROM THE VALVE DISLODGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263125 | ENVEO PRO DELIVERY SYSTEM | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV | NPT | MEDTRONIC HEART VALVES DIVISION | ENVPRO-16-US | 0012138225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Female | Required Intervention | "SEE H11...." |