ARCHITECT CMV IGG REAGENT KIT
Report
- Report Number
- 3008344661-2024-00093
- Event Type
- Malfunction
- Date Received
- August 14, 2024
- Date of Event
- July 25, 2024
- Report Date
- September 30, 2024
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- LFZ
- UDI-DI
- 0038074021117
- PMA / PMN Number
- K220949
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION HAS BEEN INCLUDED. NO ADDITIONAL PATIENT DETAILS ARE AVAILABLE.
DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. TICKET SEARCH BY LOT INDICATES NORMAL COMPLAINT ACTIVITY. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE CUSTOMER¿S OBSERVATION. A REVIEW OF TICKET TRENDING DID NOT IDENTIFY ANY RELATED TREND REGARDING COMMONALITIES FOR COMPLAINT LOT NUMBER AND ISSUE. CLINICAL SENSITIVITY TESTING MET ACCEPTANCE CRITERIA AND THE PRODUCT IS PERFORMING AS EXPECTED. LABELLING REVIEW CONCLUDES THAT THE ISSUE IS ADEQUATELY ADDRESSED. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT CMV IGG ASSAY WAS IDENTIFIED.
THE CUSTOMER OBSERVED FALSE NONREACTIVE ARCHITECT CMV IGG RESULTS FOR ONE PREGNANT PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2024 IGG NEGATIVE; IGM POSITIVE: 2.95 S/CO. (B)(6) 2024 IGG NEGATIVE, IGM POSITIVE: 2.51 S/CO. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSE NONREACTIVE ARCHITECT CMV IGG RESULTS FOR ONE PREGNANT PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2024 IGG NEGATIVE; IGM POSITIVE: 2.95 S/CO. (B)(6) 2024 IGG NEGATIVE, IGM POSITIVE: 2.51 S/CO. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262164 | ARCHITECT CMV IGG REAGENT KIT | ENZYME LINKED IMMUNOABSORBENT ASSAY, CYTOMEGALOVIRUS | LFZ | ABBOTT IRELAND DIAGNOSTICS DIVISION | 61380FZ00 | 0038074021117 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | ARC I2K PROC MOD, 03M74-01, (B)(6) | ARC I2K PROC MOD, 03M74-01, (B)(6) |