ANALYTICAL P MODULE
Report
- Report Number
- 1823260-2011-00978
- Event Type
- Malfunction
- Date Received
- February 22, 2011
- Date of Event
- February 3, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K953239
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE USER INTERMITTENTLY RECEIVED QUESTIONABLE CREATININE PLUS/ENZYMATIC (CREATININE) RESULTS AND PROVIDED DATA FOR TWO PATIENT SAMPLES. ALL REPEAT TESTING WAS PERFORMED ON ANOTHER ANALYTICAL P MODULE (SERIAL NUMBER WAS NOT PROVIDED). ALL RESULTS ARE IN MG/DL. PATIENT SAMPLE 1 INITIAL RESULT WAS 1.7. THE REPEAT RESULT WAS 1.7, AND ON (B)(6) 2011 WERE 0.9 AND 0.9. THE SAME SAMPLE FROM THE PATIENT HAD BEEN TESTED ON A POINT OF CARE (POC) DEVICE WHILE IN THE ER PRIOR TO THE TESTING OF THE SAMPLE IN THE LABORATORY. THE RESULT FROM THE POC DEVICE WAS 0.6. PATIENT SAMPLE 2 INITIAL RESULT WAS 1.7. THE REPEAT RESULT WAS 1.9, AND ON THE NEXT DAY WERE 0.9 AND 0.9. THE INITIAL RESULT WAS FLAGGED WITH A DELTA CHECK IN THE HOST SYSTEM TRIGGERING THE REPEAT TESTING. THE DATE OF TESTING FOR PATIENT SAMPLE 2 WAS UNKNOWN. THE ERRONEOUS RESULTS WERE REPORTED OUTSIDE THE LABORATORY. NEITHER PATIENT WAS TREATED OR AFFECTED DUE TO THE ERRONEOUS RESULT. THE CREATININE REAGENT LOT NUMBER WAS NOT PROVIDED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THERE WAS A LOOSE NUT CONNECTION TO THE REAGENT SYRINGE. HE PERFORMED PREVENTIVE MAINTENANCE AND TIGHTENED THE CONNECTIONS TO THE SYRINGES. TO VERIFY THE ANALYZER OPERATION, HE TESTED THE SYSTEM IN DIAGNOSTIC MODE. THE USER RAN QUALITY CONTROL WITH ACCEPTABLE RESULTS AND RAN PRECISION TESTING WITH RESULTS WITHIN SPECIFICATIONS. THE FIELD APPLICATION SPECIALIST ADJUSTED THE REAGENT PROBES AND REPLACED THE REAGENT STIRRER AS IT WAS BENT. TO VERIFY THE ANALYZER OPERATION, HE RAN PRECISION TESTING THEN RELOADED THE ORIGINAL REAGENT PACK AND RAN PRECISION TESTING. BOTH PRECISIONS WERE ACCEPTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANALYTICAL P MODULE | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |