TRIAGE CARDIAC PROFILER KIT
Report
- Report Number
- 2027969-2011-00289
- Event Type
- Injury
- Date Received
- February 9, 2011
- Date of Event
- January 18, 2011
- Report Date
- February 9, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- NBC
- PMA / PMN Number
- K030286
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
PRODUCT SUPPORT COULD NOT RULE OUT SAMPLE SPECIFIC INTERFERENCE WITHOUT RETURN OF PT SAMPLE. PROFILER LOT W48339 HAS BEEN PREVIOUSLY TESTED. NO DISCREPANT RESULTS WERE OBSERVED WITH RETAIN DEVICES. CONCLUSION: PRODUCT SUPPORT TESTED IN HOUSE CAL H(POS CTRL) AND CAL Z (NEG CTRL) ON RETAIN LOT #W48339. RESULTS FOR TNI RECOVERY ARE BELOW: CAL H: ALL 3 REPLICATES TNI RESULTS WERE WITHIN THE MFG 1SD RANGE. CAL Z: PRODUCT SUPPORT OBSERVED ALL DATA POINTS FOR CAL Z TO BE BELOW THE MFG CUT OFF OF 0.10NG/ML. UNABLE TO DETERMINE CAUSE OF CUSTOMERS POSITIVE RESULT WITHOUT SAMPLE. NO PRODUCT DEFICIENCY WAS ESTABLISHED. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER REPORTED A FALSE POSITIVE TROPONIN (TNI) RESULT USING THE TRIAGE CARDIAC PROFILER. PT WENT TO THE ED COMPLAINING OF CHEST PAINS. PT WAS ADMITTED AND LATER DISCHARGED. DIAGNOSIS: RHABDOMYOLYSIS, ACUTE BRONCHITIS. TRIAGE TNI CUT OFF: 0-0.4. DIMENSION LAB ANALYZER: 0-0.5. SERIAL DRAWS WERE TESTED ON LAB ANALYZER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIAGE CARDIAC PROFILER KIT | CARDIAC MARKER TEST | NBC | ALERE SAN DIEGO, INC. | 97100CP | W48339 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization |