FDA Adverse Event Injury Summary report: N

TRIAGE CARDIAC PROFILER KIT

MDR report key: 1997515 · Received February 9, 2011

Report

Report Number
2027969-2011-00289
Event Type
Injury
Date Received
February 9, 2011
Date of Event
January 18, 2011
Report Date
February 9, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
NBC
PMA / PMN Number
K030286
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT SUPPORT COULD NOT RULE OUT SAMPLE SPECIFIC INTERFERENCE WITHOUT RETURN OF PT SAMPLE. PROFILER LOT W48339 HAS BEEN PREVIOUSLY TESTED. NO DISCREPANT RESULTS WERE OBSERVED WITH RETAIN DEVICES. CONCLUSION: PRODUCT SUPPORT TESTED IN HOUSE CAL H(POS CTRL) AND CAL Z (NEG CTRL) ON RETAIN LOT #W48339. RESULTS FOR TNI RECOVERY ARE BELOW: CAL H: ALL 3 REPLICATES TNI RESULTS WERE WITHIN THE MFG 1SD RANGE. CAL Z: PRODUCT SUPPORT OBSERVED ALL DATA POINTS FOR CAL Z TO BE BELOW THE MFG CUT OFF OF 0.10NG/ML. UNABLE TO DETERMINE CAUSE OF CUSTOMERS POSITIVE RESULT WITHOUT SAMPLE. NO PRODUCT DEFICIENCY WAS ESTABLISHED. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER REPORTED A FALSE POSITIVE TROPONIN (TNI) RESULT USING THE TRIAGE CARDIAC PROFILER. PT WENT TO THE ED COMPLAINING OF CHEST PAINS. PT WAS ADMITTED AND LATER DISCHARGED. DIAGNOSIS: RHABDOMYOLYSIS, ACUTE BRONCHITIS. TRIAGE TNI CUT OFF: 0-0.4. DIMENSION LAB ANALYZER: 0-0.5. SERIAL DRAWS WERE TESTED ON LAB ANALYZER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIAGE CARDIAC PROFILER KIT CARDIAC MARKER TEST NBC ALERE SAN DIEGO, INC. 97100CP W48339

Patients

Seq Age Sex Outcome Treatment
1 48 YR Hospitalization