EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM
Report
- Report Number
- 2015691-2024-06080
- Event Type
- Injury
- Date Received
- August 13, 2024
- Date of Event
- July 1, 2024
- Report Date
- September 11, 2024
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPW
- UDI-DI
- 00690103212846
- PMA / PMN Number
- P230013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED/ADDED: B4, G3, G6, H2, H6 AND H11. THE COMPLAINT FOR VALVE INTERACTS WITH CONDUCTION SYSTEM WAS CONFIRMED WITH OTHER EMPIRICAL EVIDENCE VIA INFORMATION REPORTED BY EDWARDS CLINICAL SPECIALIST. NO IMAGERY WAS MADE AVAILABLE FOR EVALUATION. PER THE INSTRUCTIONS FOR USE (IFU), CONDUCTION SYSTEM DISTURBANCE AND/OR HEART BLOCK WHICH MAY REQUIRE TREATMENT (INCLUDING PERMANENT PACEMAKER) ARE POTENTIAL ADVERSE EVENTS. HEART BLOCK AND OTHER CONDUCTION DISTURBANCES ARE COMMON IN PATIENTS WITH UNDERLYING CARDIOVASCULAR DISEASE AND CAN BE EXACERBATED OR AGGRAVATED WITH STANDARD INTRA-OPERATIVE MEDICATIONS OR ANESTHESIA. SUCH EVENTS ARE RELATED TO THE PATIENT'S UNDERLYING CONDITION. OTHER MECHANISMS FOR CONDUCTION ABNORMALITIES COULD BE RELATED TO CORONARY OBSTRUCTION DUE TO AIR EMBOLISM, CORONARY SPASM AND/OR LOCAL EDEMA AFFECTING THE AV NODE. CONDUCTION DISTURBANCES HAVE ALSO BEEN DOCUMENTED DURING TRANSCATHETER CARDIAC PROCEDURES AS A RESULT OF DIRECT INTERACTION WITH CARDIAC ANATOMY, ESPECIALLY IN PATIENTS WITH UNDERLYING CONDUCTION ABNORMALITIES. OTHER POTENTIAL CAUSES INCLUDE TRAUMA BY A GUIDEWIRE OR DELIVERY SYSTEM. HOWEVER, A DEFINITE ROOT CAUSE WAS UNABLE TO BE DETERMINED FOR THIS EVENT.
THE EVENT IS CAPTURED BY EDWARDS LIFESCIENCES UNDER COMPLAINT #: (B)(4). THE MANUFACTURER'S INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
EDWARDS RECEIVED NOTIFICATION OF AN EVOQUE VALVE IN TRICUSPID POSITION WHERE ON POST-OPERATIVE DAY (POD) 4, PATIENT DEVELOPED CONTINUOUS AV-BLOCK. A LEADLESS PACEMAKER WAS IMPLANTED AND THE PATIENT WAS NOTED TO BE IN STABLE CONDITION AFTERWARDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1894690 | EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM | PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES | NPW | EDWARDS LIFESCIENCES | 9850EV52 | 00690103212846 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |