FDA Adverse Event Injury Summary report: N

EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM

MDR report key: 19974428 · Received August 13, 2024

Report

Report Number
2015691-2024-06080
Event Type
Injury
Date Received
August 13, 2024
Date of Event
July 1, 2024
Report Date
September 11, 2024
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPW
UDI-DI
00690103212846
PMA / PMN Number
P230013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED/ADDED: B4, G3, G6, H2, H6 AND H11. THE COMPLAINT FOR VALVE INTERACTS WITH CONDUCTION SYSTEM WAS CONFIRMED WITH OTHER EMPIRICAL EVIDENCE VIA INFORMATION REPORTED BY EDWARDS CLINICAL SPECIALIST. NO IMAGERY WAS MADE AVAILABLE FOR EVALUATION. PER THE INSTRUCTIONS FOR USE (IFU), CONDUCTION SYSTEM DISTURBANCE AND/OR HEART BLOCK WHICH MAY REQUIRE TREATMENT (INCLUDING PERMANENT PACEMAKER) ARE POTENTIAL ADVERSE EVENTS. HEART BLOCK AND OTHER CONDUCTION DISTURBANCES ARE COMMON IN PATIENTS WITH UNDERLYING CARDIOVASCULAR DISEASE AND CAN BE EXACERBATED OR AGGRAVATED WITH STANDARD INTRA-OPERATIVE MEDICATIONS OR ANESTHESIA. SUCH EVENTS ARE RELATED TO THE PATIENT'S UNDERLYING CONDITION. OTHER MECHANISMS FOR CONDUCTION ABNORMALITIES COULD BE RELATED TO CORONARY OBSTRUCTION DUE TO AIR EMBOLISM, CORONARY SPASM AND/OR LOCAL EDEMA AFFECTING THE AV NODE. CONDUCTION DISTURBANCES HAVE ALSO BEEN DOCUMENTED DURING TRANSCATHETER CARDIAC PROCEDURES AS A RESULT OF DIRECT INTERACTION WITH CARDIAC ANATOMY, ESPECIALLY IN PATIENTS WITH UNDERLYING CONDUCTION ABNORMALITIES. OTHER POTENTIAL CAUSES INCLUDE TRAUMA BY A GUIDEWIRE OR DELIVERY SYSTEM. HOWEVER, A DEFINITE ROOT CAUSE WAS UNABLE TO BE DETERMINED FOR THIS EVENT.

Additional Manufacturer Narrative · 0

THE EVENT IS CAPTURED BY EDWARDS LIFESCIENCES UNDER COMPLAINT #: (B)(4). THE MANUFACTURER'S INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

EDWARDS RECEIVED NOTIFICATION OF AN EVOQUE VALVE IN TRICUSPID POSITION WHERE ON POST-OPERATIVE DAY (POD) 4, PATIENT DEVELOPED CONTINUOUS AV-BLOCK. A LEADLESS PACEMAKER WAS IMPLANTED AND THE PATIENT WAS NOTED TO BE IN STABLE CONDITION AFTERWARDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1894690 EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES NPW EDWARDS LIFESCIENCES 9850EV52 00690103212846

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention