FDA Adverse Event Malfunction Summary report: N

ALINITY I FREE T4 REAGENT KIT

MDR report key: 19974147 · Received August 13, 2024

Report

Report Number
3005094123-2024-00417
Event Type
Malfunction
Date Received
August 13, 2024
Date of Event
August 6, 2024
Report Date
November 12, 2024
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
CEC
UDI-DI
00380740121822
PMA / PMN Number
K173122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY I FREE T4 RESULTS INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND IN HOUSE TESTING. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. RETURN TESTING WAS NOT PERFORMED AS RETURNS WERE NOT AVAILABLE. AN INCREASE IN COMPLAINTS HAS BEEN OBSERVED FOR LOT 63089UD00, HOWEVER, IN-HOUSE PERFORMANCE WAS COMPLETED WHICH INDICATES THE PRODUCT IS PERFORMING AS EXPECTED. THE DEVICE HISTORY RECORD REVIEW DID NOT SHOW ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES, OR DEVIATIONS ASSOCIATED WITH THE LOT NUMBER 63089UD00 AND COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I FREE T4 FOR LOT NUMBER 63089UD00 WAS IDENTIFIED. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE ELEVATED ALINITY I FREE T4 RESULTS FOR ONE SAMPLE. (CUSTOMER PROVIDED RANGE 8-20 PMOL/L). (B)(6) 2024 SID (B)(6) INITIAL RESULT = > 64PMOL/L, REPEAT RESULT ON ANOTHER ALINITY = 14.4 PMOL/L. REPEATED (SN (B)(6)) RESULTS = 14.0 PMOL/L, 13.9 PMOL/L, 14.4 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE ELEVATED ALINITY I FREE T4 RESULTS FOR ONE SAMPLE. (CUSTOMER PROVIDED RANGE 8-20 PMOL/L). (B)(6) 2024 SID (B)(6) INITIAL RESULT = > 64PMOL/L, REPEAT RESULT ON ANOTHER ALINITY = 14.4 PMOL/L REPEATED (SN (B)(6)) RESULTS = 14.0 PMOL/L, 13.9 PMOL/L, 14.4 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1176249 ALINITY I FREE T4 REAGENT KIT RADIOIMMUNOASSAY, FREE THYROXINE CEC ABBOTT IRELAND DIAGNOSTICS DIVISION 63089UD00 00380740121822

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown