FDA Adverse Event Malfunction Summary report: N

ACUSON SC2000 ULTRASOUND SYSTEM

MDR report key: 19973464 · Received August 13, 2024

Report

Report Number
3023245-2024-00056
Event Type
Malfunction
Date Received
August 13, 2024
Date of Event
August 8, 2024
Report Date
September 27, 2024
Manufacturer
SIEMENS HEALTHINEERS
Product Code
IYN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION IS ONGOING AT SIEMENS HEALTHINEERS TO IDENTIFY THE ROOT CAUSE OF THE ISSUE. THE REPORT WILL BE SUPPLEMENTED AFTER THE INVESTIGATION IS COMPLETED. REFERENCE# (B)(4).

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE INVESTIGATION INFORMATION. IN THIS CASE, THE ROOT CAUSE OF THIS ISSUE WAS DETERMINED TO RESULT FROM A SOFTWARE EXCEPTION ERROR DURING THE STRESS ECHO EXAM. THE PROBABLE REASON FOR THE ERROR IS DUE TO THE STRESS ECHO SUBSYSTEM ATTEMPTING TO ACCESS A BACKUP FILE BEING CONTROLLED BY SOLIDCORE AT THE SAME TIME. SOFTWARE WAS RELOADED ON THE SYSTEM TO RESOLVE THE ISSUE. THE OCCURRENCE OF THIS ERROR WAS RESOLVED VIA A FIX IN SOFTWARE. REFERENCE#: (B)(4).

Description of Event or Problem · 0

WHILE USING THE SC2000 FOR A DOBUTAMINE STRESS TEST, AN ERROR MESSAGE APPEARED. THIS OCCURRED DURING POST IMAGING. AFTER THE ERROR MESSAGE POPPED UP, THE CUSTOMER WAS UNABLE TO REVIEW ALL THE IMAGES FROM THE STRESS TEST STAGES INCLUDING BASELINE, FIRST STAGE, AND POST INFUSION. THEY DECIDED TO END THE STUDY, HOPING THE IMAGES WOULD BE PRESERVED AND TRANSFERRED TO THE PACS SYSTEM. HOWEVER, THIS DID NOT HAPPEN AND DATA WAS LOST. THE FILE ONLY CONTAINED IMAGES CAPTURED PRIOR TO STARTING THE STRESS TEST. THE EXAMINATION HAD TO BE REPEATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1639271 ACUSON SC2000 ULTRASOUND SYSTEM ULTRASOUND DEVICE IYN SIEMENS HEALTHINEERS 10433816

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other| H