FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER.

MDR report key: 1997109 · Received February 21, 2011

Report

Report Number
2050012-2011-00515
Event Type
Malfunction
Date Received
February 21, 2011
Date of Event
January 23, 2011
Report Date
January 23, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER STATED THAT THERE IS A BROKEN Y FITTING IN HYDRO AND NO FOAM IS LEAKING FROM IT. SERVICE WAS NOT DISPATCHED. A FIELD SERVICE ENGINEER (FSE) SENT THE PART TO THE CUSTOMER TO REPLACE THE Y FILLING. FSE VERIFIED CUSTOMER HAD REPLACED THE Y FITTING AND NO LEAKS WERE NOTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO NO FOAM SOLUTION LEAK. NO INJURY OR EXPOSURE WAS REPORTED. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER. CLINICAL CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. DXC600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1