NA
Report
- Report Number
- 3006630150-2024-05264
- Event Type
- Malfunction
- Date Received
- August 12, 2024
- Date of Event
- July 17, 2024
- Report Date
- December 31, 2024
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
- Product Code
- GXD
- UDI-DI
- 00813250010107
- PMA / PMN Number
- K050084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT.
BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. BLOCK 26B: GXD, GXI.
DEVICE ANALYSIS PERFORMED ON THE RETURNED ELECTRODE REVEALED NORMAL DEVICE CHARACTERISTICS DURING VISUAL AND FUNCTIONAL TESTING. THEREFORE, WITH ALL THE AVAILABLE INFORMATION, ENGINEERS CONCLUDED THAT THE CAUSE OF THE REPORTED EVENT WAS UNABLE TO BE DETERMINED. B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. 26B: GXD, GXI.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RADIOFREQUENCY ABLATION (RFA) PROCEDURE DURING WHICH AN ELECTRODE BECAME DETACHED FROM THE CONNECTOR WHILE INSIDE THE PATIENT. THE OUTCOME OF THE PROCEDURE WAS UNABLE TO BE OBTAINED. AFTER FURTHER REVIEW, THE BROKEN ELECTRODE WAS DISCOVERED POST-PROCEDURE AND DID NOT OCCUR INSIDE THE PATIENT. AS SUCH, THERE WAS NO REPORTED PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2174576 | NA | GENERATOR, LESION, RADIOFREQUENCY | GXD | BOSTON SCIENTIFIC NEUROMODULATION CORPORATION | CSK-TC10 | 31746134 | 00813250010107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |