OMNIPOD 5 POD
Report
- Report Number
- 3004464228-2024-31551
- Event Type
- Malfunction
- Date Received
- August 12, 2024
- Date of Event
- July 27, 2024
- Report Date
- August 12, 2024
- Manufacturer
- INSULET CORPORATION
- Product Code
- QFG
- UDI-DI
- 10385083000114
- PMA / PMN Number
- K203768
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. IT WAS REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. THIS CONDITION COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.
IT WAS REPORTED THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO OVER 250 MG/DL WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. IN ADDITION, THE PATIENT REPORTS THE PODS ADHESIVE HAD SEPARATED FROM THE POD INFUSION SITE (ABDOMEN), CAUSING THE CANNULA TO BECOME DISLODGED. AS TREATMENT, THE PATIENT APPLIED A NEW P OD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2028358 | OMNIPOD 5 POD | ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP | QFG | INSULET CORPORATION | PT-000435 | PH1K04182321 | 10385083000114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Male |