FDA Adverse Event Summary report: N

UNICEL® DXC 880I SYNCHRON® ACCESS® CLINICAL SYSTEM (FULL SYSTEM)

MDR report key: 1995091 · Received February 17, 2011

Report

Report Number
2050012-2011-00428
Date Received
February 17, 2011
Date of Event
June 14, 2010
Report Date
December 22, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLES WERE SERUM COLLECTED IN 13 X 75 MM TUBE. THE SAMPLES WERE ALIQUOTED BY THE AUTOMATE AND TESTED ON THE DXC 800. THE SAMPLE VOLUME USED IN THE TESTING IS 3 UL. THE CUSTOMER RAN THERAPEUTIC DRUG MONITORING (TDM) CONTROLS AFTER LOADING AND CALIBRATING REAGENT ONCE A DAY. ALL THE DAILY QC RESULTS MET ESTABLISHED RANGES. A BCI FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENTS AND DID NOT NOTE OR IDENTIFY ANY ISSUES A CLEAR ROOT CAUSE IS UNKNOWN FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUS LOW GENTAMICIN RESULTS ON TWO PATIENTS GENERATED BY THE UNICEL DXC 880I SYNCHRON ACCESS CLINICAL SYSTEM. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. THE HOSPITAL'S PHARMACIST IDENTIFIED THE RESULTS AS ERRONEOUS AND NOTIFIED THE CUSTOMER. THE SAMPLES WERE REPEATED AND HIGHER RESULTS WERE OBTAINED. AMENDED REPORTS WERE INITIATED. PATIENTS' TREATMENT WAS NOT IMPACTED BY THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 880I SYNCHRON® ACCESS® CLINICAL SYSTEM (FULL SYSTEM) DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER, JJE BECKMAN COULTER INC. DXC 880I N/A

Patients

Seq Age Sex Outcome Treatment
1