FDA Adverse Event Malfunction Summary report: N

TROCLEAN, 5-7MM & 7-12MM

MDR report key: 19950690 · Received August 9, 2024

Report

Report Number
1060680-2023-00011
Event Type
Malfunction
Date Received
August 9, 2024
Date of Event
August 24, 2023
Report Date
October 27, 2023
Manufacturer
SUPER BRUSH LLC
Product Code
GEA
UDI-DI
00749756912822
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A USER FACILITY REPORTED, "THE SPONGE ATTACHED TO THE STICK SEPARATED DURING USE AND HAD TO BE RETRIEVED FORM THE TROCAR/PATIENT WHICH CAUSED A DELAY IN THE PROCEDURE." DEROYAL INDUSTRIES REQUESTED RETURN OF THE SAMPLE, BUT IT WAS STATED THAT IT COULD NOT BE RETURNED. THIS INVESTIGATION IS NOT COMPLETE AT THIS TIME. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. WE WILL PROVIDE FOLLOW-UP REPORT WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

A SUPPLIER CORRECTIVE ACTION REQUEST (SCAR) WAS ISSUED TO THE SUPPLIER, SUPERBRUSH LLC. DEROYAL INVESTIGATED INVENTORY AND FOUND THAT ALL WERE WITHIN SPECIFICATION. A COMPLAINT TO SALES RATIO WAS CALCULATED FROM JUNE 2021 TO AUGUST 2023 AND FOUND TO BE (B)(4). ROOT CAUSE: BECAUSE THE SAMPLE WAS UNABLE TO BE RETURNED, THE ROOT CAUSE OF THE COMPLAINT WAS UNABLE TO BE DETERMINED. CORRECTIVE AND PREVENTIVE ACTION: AN UPDATE TO LABELING WAS MADE TO CLARIFY THE SIZES OF TROCARS TO BE USED WITH THE SMALL AND LARGE SIZE BRUSHES. DEROYAL WILL CONTINUE TO MONITOR FOR TRENDS AROUND THIS ITEM AND ISSUE. THIS INVESTIGATION IS COMPLETE AT THIS TIME. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. WE WILL PROVIDE FOLLOW-UP REPORT IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

A USER FACILITY REPORTED, "THE SPONGE ATTACHED TO THE STICK SEPARATED DURING USE AND HAD TO BE RETRIEVED FORM THE TROCAR/PATIENT WHICH CAUSED A DELAY IN THE PROCEDURE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2201074 TROCLEAN, 5-7MM & 7-12MM CANNULA, SURGICAL, GENERAL & PLASTIC SURGERY GEA SUPER BRUSH LLC 28-1002 58329415 00749756912822

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown