FDA Adverse Event Malfunction Summary report: N

NEO. TEMP PROBE 12 PINK, FOR USE WITH NI CABLE

MDR report key: 19950682 · Received August 9, 2024

Report

Report Number
9613793-2023-00016
Event Type
Malfunction
Date Received
August 9, 2024
Date of Event
July 14, 2023
Report Date
August 2, 2023
Manufacturer
DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L
Product Code
FMT
UDI-DI
00749756885447
PMA / PMN Number
K925006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A USER FACILITY REPORTED, "BABY TEMPERATURE IS READING HYPOTHERMIC AND IT IS GETTING THE BABIES TOO HOT" A SAMPLE WAS REQUESTED TO BE RETURNED TO DEROYAL INDUSTRIES BUT WAS NOT ABLE TO BE RETURNED. DEROYAL REVIEWED FAILURE MODES AND EFFECTS ANALYSIS (FMEA) AND CORRECTIVE ACTION AND PREVENTIVE ACTION AND FOUND NO ISSUES. AN INVENTORY CHECK WAS PERFORMED ON 125 PROBES AND NO ISSUES WERE FOUND. (B)(4). ROOT CAUSE: BECAUSE NO SAMPLE WAS RETURNED AND IT WAS NOT POSSIBLE TO PERFORM AN INSPECTION OF THE LOT INVOLVED, A ROOT CAUSE COULD NOT BE DETERMINED. CORRECTIVE AND PREVENTIVE ACTIONS: NO ACTIONS WERE TAKEN AS THE ROOT CAUSE COULD NOT BE IDENTIFIED. DEROYAL WILL CONTINUE TO MONITOR FOR TRENDS SURROUNDING THIS ITEM AND ISSUE. THIS INVESTIGATION IS COMPLETE AT THIS TIME. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. WE WILL PROVIDE FOLLOW-UP REPORT IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

A USER FACILITY REPORTED, "BABY TEMPERATURE IS READING HYPOTHERMIC AND IT IS GETTING THE BABIES TOO HOT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2201066 NEO. TEMP PROBE 12 PINK, FOR USE WITH NI CABLE WARMER, INFANT RADIANT FMT DEROYAL CIENTIFICA DE LATINOAMERICA S.R.L FNICU-37 00749756885447

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other