FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 19949343 · Received August 9, 2024

Report

Report Number
1000113657-2024-00335
Event Type
Malfunction
Date Received
August 9, 2024
Date of Event
July 16, 2024
Report Date
August 26, 2024
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE RETURNED - PRODUCT EVALUATION IN-PROCESS. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. NOTE: MANUFACTURER CONTACTED CUSTOMER IN SEVERAL FOLLOW-UP CALLS TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER AT THIS TIME.

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 26-AUG-2024: METER AND TEST STRIPS WERE RETURNED FOR EVALUATION. PRODUCT TESTING WAS PERFORMED AND NO DEFECT FOUND ON RETURNED METER AND TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 218, 191, 172, 215 AND 175 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT IS 110 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK-TO-BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 06/30/2025 AND TEST STRIPS WERE OPENED ONE WEEK AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 218 MG/DL , DATE: ON (B)(6) 2024, TIME: 8:33AM FASTING . RESULT 2: 191 MG/DL , DATE: ON (B)(6) 2024, TIME: 8:00AM FASTING. RESULT 3: 172 MG/DL , DATE: ON (B)(6) 2024, TIME: 8:00AM FASTING . RESULT 4: 215 MG/DL, DATE: ON (B)(6) 2024 , TIME: 8:15AM FASTING. RESULT 5: 175 MG/DL, DATE: ON (B)(6) 2024 , TIME: 8:30AM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2486218 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL ZB5500S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown