FDA Adverse Event Malfunction Summary report: N

ANGLED REAMER, DRIVE SHAFT

MDR report key: 19949152 · Received August 9, 2024

Report

Report Number
1220246-2024-07117
Event Type
Malfunction
Date Received
August 9, 2024
Date of Event
July 24, 2024
Report Date
September 27, 2024
Manufacturer
ARTHREX, INC.
Product Code
LXH
UDI-DI
00888867343436
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

COMPLAINT ALLEGATION IS CONFIRMED. ONE UNPACKAGED AR-9676, BATCH 022338, WAS RECEIVED FOR INVESTIGATION. THE REAMER ARRIVED CONTAINED WITHIN THE AR-9597-10, BUT IT WAS NOT LODGED INSIDE. UPON VISUAL INSPECTION, SCRATCH MARKS WERE OBSERVED AT BOTH THE DISTAL AND PROXIMAL ENDS OF THE DEVICE, SUGGESTING SIGNS OF FRICTION. WHEN THE DEVICE UNDERGOES FUNCTIONAL TESTING WITHIN THE AR-9597-10, IT EXPERIENCES RESISTANCE AND FRICTION. THE OBSERVED CONDITION IS MOST LIKELY THE RESULT OF OVERHEATING DURING USE CAUSED BY OVERSTRESSING A DEVICE THAT IS DAMAGED NATURALLY AND INEVITABLY AS A RESULT OF NORMAL WEAR OR AGING.

Description of Event or Problem · 0

ON (B)((6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9597-10 ANGLED REAMER SLEEVE AND AN AR-9676 ANGLED REAMER, DRIVE SHAFT WERE DIFFICULT TO TAKE APART FROM ONE ANOTHER WHEN THE AUGMENT TRAY WAS BEING PUT UP AFTER SURGERY AND SEEMS THE INNER SHAFT HAS SCUFF MARKS NOT TO ROTATE PROPERLY. NO ADVERSE EVENT OR ADDED TIME TO THE CASE. THIS WAS DISCOVERED AFTER USE, WITH NO REPORTED PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2492197 ANGLED REAMER, DRIVE SHAFT MANUAL INSTR, GENERAL SURGICAL LXH ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT 022338 00888867343436

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown