FDA Adverse Event Malfunction Summary report: N

QUANTUM WORKSTATION 12" ELITE

MDR report key: 19948748 · Received August 9, 2024

Report

Report Number
3006073153-2024-00039
Event Type
Malfunction
Date Received
August 9, 2024
Date of Event
July 10, 2024
Report Date
August 9, 2024
Manufacturer
SPECTRUM MEDICAL LTD
Product Code
DRY
UDI-DI
05060434422309
PMA / PMN Number
K202557
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

EVENTS OCCURRED DURING THE 1500 HOUR AS FOLLOWS: 1. ECMO SPECIALIST NOTICED NEGATIVE ART FLOW IN PARALLEL WITH ECMO PRIMER NOTICING BACKWARD FLOW 2. PATIENT CLAMPED OFF IMMEDIATELY - HAND CRANKING RESUMED 3. CLINICIANS STARTED TROUBLESTHOOTING A. RESET QWS - TRIED GETTING SWEEP GAS TO CONNECT TO TANKS BUT WAS RUNNING SHORT ON TUBING, SO DID NOT DO A HARD RESET FROM THE BOTTOM OF THE FRAME B. WENT INTO ADMIN MODE TO ENSURE THEY HAD COUNTERCLOCKWISE SELECTED, SWAPPED TO CLOCKWISE, UNCLAMPED, NO CHANGE. WENT BACK TO COUNTERCLOCKWISE, NO CHANGE. C. SWAPPED OUT ROLLER PUMP UNIT WHILE REMAINING ON SAME FRAME/QWS/QVM D. REGAINED FORWARD COUNTERCLOCKWISE FLOW 4. RED HERRING - THIS WAS IMMEDIATELY AFTER A RACEWAY WALK AND PRIOR TO STARTING, SPECIALIST TURNED DOWN FLOW TO "OFF", THERE WAS STILL A WHIFF GOING THROUGH SO PRIMER TAPPED THE STOP BUTTON (PRIOR TO LIFTING LID), SCREEN NOTIFICATION POPPED UP AND SPECIALIST QUICKLY TAPPED YES. THE CLINICIANS TROUBLESHOT TOGETHER AND THERE WAS NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1561401 QUANTUM WORKSTATION 12" ELITE MONITOR/CONSOLE DRY SPECTRUM MEDICAL LTD QWS12E 05060434422309

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown