FDA Adverse Event Malfunction Summary report: N

R-PILOT, 6X, STERILE

MDR report key: 19947852 · Received August 9, 2024

Report

Report Number
9611053-2024-00131
Event Type
Malfunction
Date Received
August 9, 2024
Report Date
October 10, 2024
Manufacturer
VDW GMBH
Product Code
EKS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

SUMMARY: INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHRS REVIEW (BATCH #1815192L, #1846455, #1826593 AND #1832274). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE (MOTOR SETTING NOT COMMUNICATED), OVERUSE (NUMBER OF USES NOT COMMUNICATED), EXCESSIVE WEAR, OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE).

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A R-PILOT, 6X, STERILE FILE BROKE DURING USE. THE OUTCOME OF THIS EVENT IS UNKNOWN AS OF THIS MDR. FURTHER INFORMATION REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1635266 R-PILOT, 6X, STERILE FILE, PULP CANAL, ENDODONTIC EKS VDW GMBH 437381

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown