AUTOSOFT 90
Report
- Report Number
- 3003442380-2024-19921
- Event Type
- Malfunction
- Date Received
- August 9, 2024
- Date of Event
- June 26, 2024
- Report Date
- September 10, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018129
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2024-19921. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INVESTIGATION ASSOCIATED WITH RELATED EVENT DATABASE (B)(4) HAS BEEN APPROVED AND IS COMPLETE. NO ADDITIONAL ACTION IS REQUIRED, AND THIS COMPLAINT WILL BE CLOSED. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION ACCORDING TO THE MARKET QUALITY REVIEW (OMQR) PROCEDURE. THE IDENTIFIED FOR MALFUNCTION CODE, ADHESIVE PATCH LIFTS OR DETACHES DURING USE, WAS NOT ASSOCIATED WITH A DESIGN OR MANUFACTURING-RELATED COMPLAINT ISSUE. THEREFORE, A DETAILED INVESTIGATION OR INSPECTION OF REFERENCE SAMPLES IS NOT REQUIRED. BATCH REVIEW: LOT 6003504 WAS MANUFACTURED ON 01-OCT-2023, IN MACHINE L171, WITH A TOTAL OF (B)(4). THE BATCH RECORD WAS REVIEWED TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED, AND NO ISSUES WERE FOUND. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WAS FOUND WITHIN THE DOCUMENTATION. CONCLUSION SUMMARY OF THE RELATED EVENT. DUE TO THE FOLLOWING BATCH RECORD REVIEW YIELDING NO DISCREPANCIES AND NO NON-CONFORMANCE (NC) WAS GENERATED DURING PRODUCTION. NO FURTHER ACTION IS REQUIRED FOR THIS COMPLAINT. THE ACRYLIC GLUE ON THE ADHESIVE IS NOT APPLIED BY CONVATEC AND THE ADHESIVE MATERIAL HAS BEEN MANUFACTURED AND INSPECTED IN ACCORDANCE WITH ALL SPECIFICATIONS, BATCH DOCUMENTATION AND THE REQUIREMENTS ACCORDING TO 21 CFR PART 820: QUALITY SYSTEM REGULATION AND ISO EN 13485:2016: MEDICAL DEVICES-QUALITY MANAGEMENT SYSTEMS.
INITIAL AND FINAL MDR (B)(4).- MDR DEVICE 3 OF 4.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED KINGDOM. IT WAS REPORTED THAT PATIENT FACED FOUR INFUSION SETS FELL OFF DURING USE EVENTS ON 26-JUN-2024. THE SET WAS IN USE FOR A FEW HOURS. THE BLOOD GLUCOSE LEVEL AT THE TIME OF EVENTS WAS 22MMOL AND PATIENT RESOLVED ALL THE EVENTS BY REPLACING INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1646532 | AUTOSOFT 90 | UNO INSET II 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002817 | 6003504 | 05705244018129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |