INSET
Report
- Report Number
- 3003442380-2024-20363
- Event Type
- Malfunction
- Date Received
- August 8, 2024
- Date of Event
- May 14, 2024
- Report Date
- November 6, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016682
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED BRAND NAME UNDER D1, COMMON DEVICE NAME UNDER D2, SERIAL NUMBER, MODEL NUMBER, PRIMARY UNIQUE DEVICE IDENTIFIER (UDI) NUMBER AND EXPIRATION DATE UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DEVICE HISTORY RECORD (DHR) REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003735, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003735 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 108 AND MANUFACTURED IN THE LINE INSET 7 ON 16/OCT/2023, WITH A TOTAL OF (B)(4) UNITS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) REVEALED THAT DURING OUTGOING TEST 6, ONE EXTENDED SAMPLING WAS RAISED: ONE FOR TUBE ATTACHED TO THE TAPE. ONE EXTENDED SAMPLING WAS ACCEPTED. THEREFORE, THE DEVICE HISTORY RECORD (DHR) CONFIRMS THAT ALL REQUIRED TESTS RELATED TO THE PROCESS WERE COMPLETED AND MET THE APPLICABLE REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: ONE EXTENDED WAS RAISED DURING THE PROCESS UNRELATED TO THE MALFUNCTION REPORTED, THEREFORE, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
INITIAL AND FINAL MDR (B)(4). DEVICE 5 OF 7.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED SEVEN INFUSION SET CANNULA WAS ROLLED INTO A BALL ON (B)(6) 2024 AND ON (B)(6) 2024. THE BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF EVENT. THE EVENT OCCURED WITH IN THREE HOURS OF INSERTION. THE SITE OF INSERTION WAS AT ABDOMEN. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 414949 | INSET | UNO INSET I 60/6 PINK TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001682 | 6003735 | 05705244016682 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Female |