FDA Adverse Event Malfunction Summary report: N

HERCULES 3 STAGE ESOPHAGEAL BALLOON

MDR report key: 1994306 · Received January 14, 2011

Report

Report Number
1037905-2011-00019
Event Type
Malfunction
Date Received
January 14, 2011
Report Date
December 17, 2010
Manufacturer
WILSON-COOK MEDICAL, INC.
Product Code
KNQ
PMA / PMN Number
K061937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVAL. WE COULD NOT CONDUCT A REVIEW OF THE DEVICE HISTORY RECORD OR CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE THE LOT NUMBER WAS NOT PROVIDED. AFTER A REVIEW OF THE ACCOUNT ORDERING AND SHIPPING HISTORY, THE LOT NUMBER COULD NOT BE DETERMINED. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCTS SAID TO BE INVOLVED WERE NOT RETURNED FOR EVAL. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. THE ADD'L INFO PROVIDED INDICATED THE BALLOON DILATOR DID NOT RECEIVE LUBRICATION PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE. THIS IS THE MOST LIKELY CAUSE FOR THE REPORTED OBSERVATION. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY A LUBRICATING AGENT TO THE BALLOON TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE ACCESSORY CHANNEL. THIS ACTIVITY WILL AID IN ENDOSCOPIC ADVANCEMENT AND BALLOON PRESERVATION. A BALLOON RUPTURE CAN OCCUR IF THE BALLOON IS INFLATED PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE OR IF THE BALLOON IS INFLATED WHILE PARTIALLY OR FULLY INSIDE THE ACCESSORY CHANNEL OF THE ENDOSCOPE. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING PRECAUTION: "DO NOT PRE-INFLATE THE BALLOON." THE INSTRUCTIONS FOR USE DIRECT THE USER TO ENSURE THE BALLOON IS COMPLETELY VISUALIZED AND POSITIONED BEFORE INFLATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING WARNING: "DURING DILATION, DO NOT INFLATE BALLOON BEYOND THE MAXIMUM INDICATED INFLATION PRESSURE." OVER INFLATION CAN CAUSE DAMAGE TO THE BALLOON DILATOR, SUCH AS A RUPTURE OR SPLIT. ANOTHER POSSIBLE CONTRIBUTING FACTOR IS USING A COMPROMISED INFLATION DEVICE TO INFLATE THE BALLOON. IF THE PRESSURE READING OF THE INFLATION DEVICE IS INACCURATE, THIS COULD CONTRIBUTE TO OVER INFLATION, POSSIBLY RESULTING IN A RUPTURE OF THE BALLOON MATERIAL. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE ESOPHAGEAL BALLOON DILATORS ARE SUBJECTED TO A VISUAL EXAMINATION TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MFG PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. CUSTOMER QA WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK ENDOSCOPY HERCULES 3 STAGE ESOPHAGEAL BALLOON. THE USER REPORTED THAT BALLOON DILATOR RUPTURED IN THE ESOPHAGUS WHEN INFLATED TO 4.5 ATM. THE DEVICE WAS REMOVED AND ANOTHER HERCULES 3 STAGE ESOPHAGEAL BALLOON WAS USED TO COMPLETE THE PROCEDURE. IT WAS CONFIRMED THAT A FOREIGN OBJECT DID NOT HAVE TO BE RETRIEVED FROM THE PT. NO ADD'L MEDICAL PROCEDURES WERE REQUIRED DUE TO THIS OCCURRENCE. THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OBSERVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HERCULES 3 STAGE ESOPHAGEAL BALLOON KNQ, DILATOR, ESOPHAGEAL KNQ WILSON-COOK MEDICAL, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 COOK ENDOSCOPY DILATION SYRINGE (DS-60CC-S)| OLYMPUS UPPER ENDOSCOPE (UNK MODEL NUMBER)