AUTOMATED PD SET W/CASSETTE4 PRONG
Report
- Report Number
- 1423500-2011-02073
- Event Type
- Malfunction
- Date Received
- February 17, 2011
- Date of Event
- January 24, 2011
- Report Date
- January 24, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE; HOWEVER, PER THE COMPLAINT INFORMATION THE CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).
(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
A CUSTOMER CONTACTED GLOBAL TECHNICAL SERVICE (GTS) REGARDING A SYSTEM ERROR 2240 ALARM THAT APPEARED ON THE HOME CHOICE (HC) DURING DRAIN 4 OF 4. THE HOME PATIENT (HP) THOUGHT, SHE WAS DONE AND DISCONNECTED. THE HP DID RECONNECT. GTS ASSISTED WITH CLEARING THE ALARM. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING FOLLOW UP THE HOME PATIENT'S (HP) NURSE SAID, SHE WAS NOT AWARE OF THE ALARM. THE MEANING OF THE ALARM WAS EXPLAINED TO THE NURSE. THE NURSE SAID, THE HP HAD NOT HAD ANY COMPLICATIONS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOMATED PD SET W/CASSETTE4 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | HOMECHOICE CYCLER |