FDA Adverse Event Malfunction Summary report: N

AU403-02E CLINICAL CHEMISTRY ANALYZER

MDR report key: 1994175 · Received February 16, 2011

Report

Report Number
2050012-2011-00473
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
January 19, 2011
Report Date
January 19, 2011
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K981743
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS DISPATCHED FOR THIS EVENT. THE AU400 WAS NOT THE SOURCE OF BURNING SMELL. A CIRCUIT CARD ON THE UPS CONNECTED TO AU400 HAD BURNT. THE FIELD SERVICE ENGINEER (FSE) PLUGGED THE AU400 TO THE WALL POWER AND VERIFIED THE INSTRUMENT OPERATION. THERE WAS NO DAMAGE TO THE AU400 AND THE OPERATION WAS NOT AFFECTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO SMOKE ODOR FROM AU400 CHEMISTRY ANALYZER. THE CUSTOMER STOPPED THE ANALYZER AND CALLED THE FIRE DEPARTMENT. THERE WAS NO EFFECT TO PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU403-02E CLINICAL CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. NA NA

Patients

Seq Age Sex Outcome Treatment
1