NONVENTED SPIKE WITH CLEARLINK
Report
- Report Number
- 6000001-2011-01077
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Date of Event
- February 1, 2011
- Report Date
- February 2, 2011
- Manufacturer
- BAXTER HEALTHCARE - MALTA
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION ON 02/09/2011. A COMPANION SAMPLE WAS SUBMITTED FOR EVALUATION FOR THE REPORTED CONDITION OF A BLOCKED CLEARLINK VALVE. THE SET WAS TESTED FOR BLOCKAGE USING A LUER-LOCKING SYRINGE. THIS ANALYSIS SHOWED THE CLEARLINK VALVE WAS BLOCKED AS DESCRIBED BY THE CUSTOMER. THE REPORTED CONDITION WAS CONFIRMED. THE ROOT CAUSE OF THIS CONDITION WAS ATTRIBUTED TO A RAW MATERIAL ISSUE (SUPPLIER ERROR/MOULDING ISSUE). A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT.
(B)(4). A COMPANION SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (U.S.); THEREFORE, IT DOES NOT HAVE A U.S. 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S.
THE CUSTOMER REPORTED TO BAXTER (B)(4) A NON-VENTED SPIKE WITH CLEARLINK SOLUTION ADMINISTRATION SET THAT WAS INVOLVED IN A NO FLOW/RESTRICTED FLOW. ACCORDING TO THE REPORT, THE NURSE STATED THE CLEARLINK VALVE IS BLOCKED BECAUSE WHEN THE OPERATOR TRIED TO INJECT OR SUCK OUT A SOLUTION THROUGH IT INTO A BAG USING A LUER-LOCK SYRINGE, THERE WAS A RESISTANCE AND THE VALVE SEEMS TO BE CLOSED. THE CONDITION WAS REPORTED TO HAVE OCCURRED DURING USE; HOWEVER, THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. THIS IS REPORT 1 OF 10 OF THE SAME REPORTED PROBLEM FROM THIS FACILITY. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NONVENTED SPIKE WITH CLEARLINK | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - MALTA | 10G03V034M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN LUER LOCKING SYRINGE |