FDA Adverse Event Malfunction Summary report: N

INNER SHEATH, FOR 26 FR. OUTER SHEATH

MDR report key: 19940615 · Received August 8, 2024

Report

Report Number
9610773-2024-01921
Event Type
Malfunction
Date Received
August 8, 2024
Date of Event
August 1, 2024
Report Date
August 27, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HIH
UDI-DI
04042761029339
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

G4 PMA/510(K) NUMBER: K931994/K931995. THE EVALUATION OF THE EVENT IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, SINCE THE DEVICE WAS NOT RETURNED FOR EVALUATION, THE DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. HOWEVER, THE LIKELY CAUSE OF THE REPORTED EVENT IS DUE TO CERAMIC TIP WAS REPAIRED BY THE THIRD PARTY WITH BAD IMPROPER ADHESIVES. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR

Description of Event or Problem · 0

IT WAS REPORTED, THE RESECTION SHEATH CERAMIC TIP WAS DEBONDING. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1069517 INNER SHEATH, FOR 26 FR. OUTER SHEATH RESECTION SHEATH HIH OLYMPUS WINTER & IBE GMBH A22040A 21102 04042761029339

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown OTV-S300 (VISERA ELITE II VIDEO SYSTEM CENTER).