ENDOPATH** XCEL* TROCAR
Report
- Report Number
- 3005075853-2011-00674
- Event Type
- Malfunction
- Date Received
- February 16, 2011
- Date of Event
- November 29, 2010
- Report Date
- November 29, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON EVALUATION, A TEST 5MM DEVICE WAS INSERTED AND REMOVED THROUGH THE TROCAR WITHOUT ANY ANOMALIES NOTED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE INCIDENT REPORTED. IN ADDITION, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT; HOWEVER, THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE DEVICE DOES NOT HAVE AN OBTURATOR. THE OBTURATOR IS THE PIECE OF THE TROCAR THAT HAS THE BATCH STAMPED. NO BATCH NUMBER WAS AVAILABLE; THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCES.
IT WAS REPORTED THAT DURING THE IMPLANTATION OF THE LAPAROSCOPIC BAND PLACEMENT, UPON REMOVAL OF THE DISSECTOR THE SURGEON WITHDREW THE DEVICE THROUGH THE TROCAR AND FELT RESISTANCE. AFTER TWO ATTEMPTS THE SURGEON REMOVED BOTH THE 5MM TROCAR AND THE DISSECTOR AS ONE UNIT AND THE TROCAR IS STILL STUCK ON THE DISSECTOR. ANOTHER 5MM TROCAR WAS USED TO COMPLETE THE CASE. THE DISSECTOR WAS NO LONGER NEEDED SINCE THE BAND WAS ALREADY PASSED. THIS WAS THE FIRST BAND PLACEMENT FOR THE PATIENT. THE DISSECTOR AND TROCAR ARE STUCK TOGETHER AND WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL* TROCAR | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DISSECTOR |