FDA Adverse Event Malfunction Summary report: N

ETS-FLEX - ENDOSCOPIC

MDR report key: 1994012 · Received February 16, 2011

Report

Report Number
3005075853-2011-00673
Event Type
Malfunction
Date Received
February 16, 2011
Report Date
January 24, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATW35 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS, THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP DONOR NEPHRECTOMY PROCEDURE, IT WAS REPORTED THAT THE STAPLER FIRED BUT THE STAPLES DID NOT CLOSE COMPLETELY, SO THERE WAS ARTERIAL BLEEDING FROM THE ARTERIAL END. THE CONSULTANT SAFELY USED CLIPS TO CONTROL THIS. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS-FLEX - ENDOSCOPIC STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1