FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1994008 · Received February 16, 2011

Report

Report Number
1423500-2011-02056
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
September 19, 2010
Report Date
January 22, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SYSTEM ERROR (SE) 2240 WAS IDENTIFIED DURING A REVIEW OF A RETURNED HOMECHOICE (HC) DEVICE WITH OCCURRENCE ON (B)(4) 2010; THERE WAS NO REPORT FOR THIS ALARM CALLED IN BY THE CUSTOMER. THERE IS NO EVIDENCE THAT PROBLEMS WITH THE HC CONTRIBUTED TO SE 2240. A ROOT CAUSE WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A SYSTEM ERROR (SE) 2240 ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. THE SYSTEM ERROR OCCURRED ON (B)(6) 2010. A SE 2240 IS A NON-RECOVERABLE ALARM THAT OCCURS WHEN THE DEVICE HAS DETECTED AIR BEING DRAWN CONTINUOUSLY INTO THE DISPOSABLE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 74 YR HOMECHOICE CYCLER