FDA Adverse Event Injury Summary report: N

OT ULTRAMINI METER

MDR report key: 1993966 · Received February 16, 2011

Report

Report Number
2939301-2011-01413
Event Type
Injury
Date Received
February 16, 2011
Date of Event
February 3, 2011
Report Date
February 6, 2011
Manufacturer
LIFESCAN INC.
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(4) 2012)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER WAS FOUND TO HAVE EEPROM FAILURE AT U6. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED. THE RETAIN TEST STRIPS ALSO PASSED ALL TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K061118.

Description of Event or Problem · 1

ON (B)(6) 2011, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRAMINI METER DISPLAYED AN "ER 5" MESSAGE. ON (B)(6) 2011, THE MEDICAL SURVEILLANCE SPECIALIST (MSS) SPOKE WITH THE PATIENT TO OBTAIN AND VERIFY INFORMATION; HOWEVER, THE PATIENT DID NOT WANT TO ANSWER FOLLOW UP QUESTIONS. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON THE MORNING OF (B)(6) 2011. THE CCA WAS ADVISED THE PATIENT MANAGES HER DIABETES WITH INSULIN (TYPE/ AMOUNT NOT SPECIFIED). IT IS NOT KNOWN WHAT ACTION THE PATIENT TOOK AT THE TIME OF THE ALLEGED ISSUE. THE PATIENT DID NOT SPECIFY DEVELOPING SYMPTOMS; HOWEVER, STATED EMERGENCY MEDICAL SERVICES (EMS) WAS CONTACTED THAT SAME MORNING. THE PATIENT STATED SHE WAS ADMINISTERED INTRAVENOUS (IV) GLUCOSE. THE PATIENT DID NOT SPECIFY WHETHER ANOTHER DEVICE WAS USED FOR TESTING. AT THE TIME OF TROUBLESHOOTING, THE CCA WALKED THE PATIENT THROUGH A RETEST TO RESOLVE THE ALLEGED ISSUE. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT ALLEGEDLY RECEIVED MEDICAL INTERVENTION FROM A HEALTH CARE PROFESSIONAL (HCP) AFTER THE REPORTED ISSUE BEGAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRAMINI METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 3056305

Patients

Seq Age Sex Outcome Treatment
1 55 YR Life Threatening| R