FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1993961 · Received February 16, 2011

Report

Report Number
1423500-2011-02049
Event Type
Malfunction
Date Received
February 16, 2011
Date of Event
January 1, 2011
Report Date
January 30, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION WAS PROVIDED. THE HOME PATIENT (HP) WAS NOT CONNECTED AT THE TIME OF THE ALARM AND DID NOT RECONNECT AFTER THE ALARM OCCURED.

Additional Manufacturer Narrative · 1

(B)(4). THE LOT NUMBER WAS NOT PROVIDED; THEREFORE A BATCH REVIEW CANNOT BE CONDUCTED. THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

A HOME PATIENT (HP) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A SYSTEM ERROR (SE) 2240 ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING USE. DURING TROUBLESHOOTING IT WAS FOUND THAT A CLAMP ON AN UNUSED SUPPLY LINE WAS OPEN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP CYCLE POWER TO CLEAR THE ALARM THEN EXPLAINED A SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP AND ADVISED THE HP TO START OVER WITH NEW SUPPLIES. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL WITH THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOME CHOICE CYCLER