FDA Adverse Event Malfunction Summary report: N

OXIMAX N-595 PULSE OXIMETER

MDR report key: 1993717 · Received January 13, 2011

Report

Report Number
2936999-2011-00032
Event Type
Malfunction
Date Received
January 13, 2011
Date of Event
December 1, 2010
Report Date
December 15, 2010
Manufacturer
COVIDIEN, FORMERLY TYCOHEALTHCARE
Product Code
DQA
PMA / PMN Number
K012891
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVAL. THE CUSTOMER REVEALED THAT THE FAILURE WAS ISOLATED TO THE SPEAKER BY SWAPPING WITH A KNOWN GOOD SPEAKER ASSEMBLY. NO FURTHER CONCLUSION CAN BE DRAWN BY THE MFG SITE SINCE THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. INFO HAS BEEN ADDED TO THE DATABASE FOR TRENDING PURPOSES.

Description of Event or Problem · 1

THE COMPANY RECEIVED A REPORT FROM A BIOMEDICAL ENGINEER THAT THE UNIT DID NOT PROVIDE AN AUDIO TONE. NO PT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OXIMAX N-595 PULSE OXIMETER PULSE OXIMETER DQA COVIDIEN, FORMERLY TYCOHEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1