FLEXLAB
Report
- Report Number
- 3010825766-2024-00003
- Event Type
- Malfunction
- Date Received
- August 8, 2024
- Date of Event
- July 1, 2024
- Report Date
- August 8, 2024
- Manufacturer
- INPECO SA
- Product Code
- CEM
- UDI-DI
- 07640172340004
- PMA / PMN Number
- K121012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ACCORDING TO THE ANALYSIS, THE RESUSPENSION OF THE CORPUSCULAR CELLS IS NOT DUE TO THE ABRUPT ASPIRATION PHASE, BUT DUE TO THE PRE-MIXING OPTION, AVAILABLE BY DEFAULT ON THE PIPETTOR OF THE ALIQUOTER MODULE. THE ALIQUOTS ARE TESTED ON A THIRD PARTY INSTRUMENT FOR COAGULATION TESTS. THE RESULTS COULD BE ALTERED BY THE PRESENCE OF CELLS (E.G. PLATELETS) IMPACTING THE PRECISION OF THE TEST. THE ROOT CAUSE OF THE EVENT IS THE PRE-MIXING OPTION OF THE ALIQUOTER MODULE'S PIPETTOR. THE PRE-MIXING OPTION HAS BEEN DISABLED IN THE NEW FIRMWARE VERSION OF THE ALIQUOTER MODULE, IN ORDER TO AVOID THE PRE-MIXING STEP BEFORE THE ASPIRATION AND DISPENSING PROCESS. THE NEW FIRMWARE HAS BEEN RELEASED BY INPECO, AND INSTALLED IN THE LABORATORY IMPACTED BY THE ISSUE. THE LABORATORY IMPACTED BY THE ISSUE IS THE ONLY CUSTOMER HAVING INSTALLED THE ALIQUOTER MODULE WITH THE FIRMWARE VERSION AFFECTED BY THE PROBLEM, AND USING THE ALIQUOTER MODULE FOR DIAGNOSTIC PURPOSES.
THE CUSTOMER REPORTED THAT THE ASPIRATION OF THE ROBOT OF THE ALIQUOTER MODULE OF FLEXLAB IS QUITE ABRUPT, AND CAUSES TURBULENCE IN THE SAMPLE TUBE.THE CITRATE TUBES FOR THE COAGULATION DON'T HAVE A GEL LAYER; THEREFORE THE CENTRIFUGED CELLS CAN BE RESUSPENDED IF THE ASPIRATION FORCES ARE TOO STRONG.THE CONSEQUENCE IS THE CREATION OF ALIQUOTS CONTAINING CORPUSCOLAR PARTICLES, THAT COULD GIVE INCORRECT RESULTS, AS THE CELLS ARE MEASURED ALONG WITH THE PLASMA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2395120 | FLEXLAB | LABORATORY AUTOMATION SYSTEM | CEM | INPECO SA | 07640172340004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |