MINMED QUICK SET
Report
- Report Number
- 3003442380-2024-19809
- Event Type
- Malfunction
- Date Received
- August 8, 2024
- Date of Event
- July 2, 2024
- Report Date
- September 10, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244017450
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4). E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2024-19809. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE COMPLAINT PR. (B)(4) HAS BEEN EVALUATED. THE LOT 6002745 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTIONS (WI) GUIDELINE FOR TEST OF REFERENCE. SAMPLES BUID-UMD VERSION 11 FOR THE LEAKAGE FROM INFUSION SITE (SPECIFIC CAUSE NOT IDENTIFIED). COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAS BEEN REQUESTED. THE FOLLOWING TEST WAS PERFORMED: VISUAL TEST ACCORDING TO WITH VERSION 7 TEST ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST 1 ACCORDING TO VERSION 10 TEST ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST 2 ACCORDING TO VERSION 25 TEST ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. PR. (B)(4) COMPLAINT TEST REPORT.PDF FOR THE DETAILS. BATCH REVIEW: THE LOT 6002745 WAS MANUFACTURED ACCORDING TO THE DOCUMENT VERSION 48 ON THE PACKING PROCESS IN THE MULTIVAC 12, ON 18/AUG/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESS HAD BEEN FULFILLED AND MET THE REQUIREMENTS. CONCLUSION SUMMARY OF THE RELATED EVENT. AS A RESULT OF THE FOLLOWING: NO REPORTED HARM, NO DEFECT ON TEST, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION. NO FURTHER ACTIONS ARE REQUIRED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN UNITED STATES. IT WAS REPORTED THAT PATIENT FACED FIVE INFUSION SETS LEAKING AT SITE EVENTS ON 02-JULY-2024. THE INFUSION SET WAS IN USE FOR 1 DAY. BLOOD GLUCOSE LEVELS REPORTED DURING THE EVENT WAS 380 MG/DL. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1530747 | MINMED QUICK SET | UNO QUICK-SET 60/6 SC1 MECA | FPA | UNOMEDICAL A/S | MMT-399A | 6002745 | 05705244017450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |