FDA Adverse Event Malfunction Summary report: N

SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS)

MDR report key: 1993612 · Received January 6, 2011

Report

Report Number
3004962788-2011-00001
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 10, 2010
Report Date
January 6, 2011
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A SITE VISIT WAS PERFORMED ON (B)(4), 2010. THE SERVICE REPORT FOR THIS VISIT STATED THAT THERE WAS A FAILURE OF THE LOCATION BOARD CABLE (LB CABLE) RESULTING FROM MISUSE. AS A RESULT, THE LB CABLE CAUSED DAMAGE TO THE LOCATION SUB SYSTEM (LSS). BOTH THE LSS AND THE LB CABLE WERE REPLACED AND THERE ARE NO FURTHER ACTIONS REQUIRED AT THIS TIME. THERE WAS NO INJURY TO THE PT REPORTED. IN AN ABUNDANCE OF CAUTION, THIS EVENT IS BEING REPORTED DUE TO THE ADDITIONAL POTENTIAL RISK ASSOCIATED WITH MULTIPLE EXPOSURES TO GENERAL ANESTHESIA.

Description of Event or Problem · 1

A SITE REPORTED THAT THEY COULD NOT GET THE PT SENSOR TRIPLETT (PST) OR LOCATABLE GUIDE (LG) TO APPEAR IN THE SENSING VOLUME, THE LIGHTS WOULDN'T TURN GREEN. THEY TRIED RESETTING THE LOCATION SUB SYSTEM (LSS), RESTARTING THE SYSTEM AND DISCONNECTING/RECONNECTING CABLES WITH NO CHANGE. THE CASE WAS CANCELLED WITH THE PT UNDER GENERAL ANESTHESIA. THERE WAS NO HARM OR INJURY TO THE PT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERDIMENSION INREACH BRONCHUS SYSTEM (SDBS) COMPUTED TOMOGRAPHY X-RAY SYSTEM JAK SUPERDIMENSION INC.

Patients

Seq Age Sex Outcome Treatment
1